Before Transfer of applications:
1) Archive the data from within the program.
> Tools > Utilities > Archive Restore.
2) Copy/Move the ArchiveRestore folder created to a network location or a cd so that it can be transferred to the new pc. ***C:Program Files Bio-Rad Laboratories Unity Real-Timel ArchiveRestore.
3) Save the Connectivity Transformation manager .txml file in the same area as the Archive. These two contain data for both the Unity Real-Time and the Transformation Manager. The .txml file is located in C:Program Files Bio-Rad Laboratories Connectivity Transformation Manager Transformer Instrument Driver- in this case it is Dimension.
During Transfer of applications:
1) The installation of the Unity Real-Time is standard.
2) The installation of the Transformation Manager is standard as long as the user has the cd sent to them from Bio-Rad.
After Transfer of applications:
1) Restore the Archive into Unity Real-Time using the Restore function.
> Tools > Utilities > Archive Restore.
2) Replace the saved .txml file into the same location on the new pc as the original pc. If a prompt appears asking to overwrite existing file, then click yes because the existing file on the new pc is blank.
*****Note: If the pc just contains the database, you will need to do a database backup instead of an archive. The .txml will not need to be moved because that information is stored on the pc with the application.
Database backup:
Go to Start Menu > Programs > Bio-Rad Laboratories > Unity Real-Time Database Utilities > Backup/Restore then choose backup and the file created will be named QCDAO.bak.
***When restoring the Database from the backup file created above, please remember to have the file located at “C:”.